Purchase of a battery cycler
Chalmers Tekniska Högskola Aktiebolag
Direct award
Request for tender:
Description of goods/services
A battery cycler.
Requirements specifications of goods/services
General requirements:
Must have at least 4 channels
Must be able to measure current, voltage and temperature on each channel
Must have protection against over-temperature, over/under-voltage and over-current of the connected cells
Must have mounting and cables that allow for channels to operate on cells inside a climate chamber
Must allow for cycling of 18650 and 21700 cylindrical cells
Should include electrical impedance spectroscopy (EIS) on each channel; in case please specify price per channel with and without EIS
Acquisition frequency
Must have a measurement acquisition frequency of at least 10 Hz
Should have a measurement acquisition frequency of at least 15 Hz
Operating range requirements
Must be able to provide a charge and discharge current range between 10mA and 20A
Should be able to provide a maximum charge and discharge current of at least 25 A
Must be able to provide a total bidirectional power flow of at least 100 W at least (20 A @ 5V)
Should be able to provide a total bidirectional power flow of at least 125 W (25 A @ 5V)
Must be able to provide a voltage range between at least 0 and 5.5V
Must have a temperature measurement range between at least -20°C and 50°C
Must have cables and connectors that ensure operating range between at least -25°C and 60°C.
Measurement precision and accuracy
Must provide a current measurement precision of at least 0.02% FSR
Should provide a current measurement precision of at least 0.01% FSR
Must provide a current measurement accuracy of at least 0.5% FSR
Must provide a current measurement accuracy of at least 0.05% below 10 A
Should provide a current measurement accuracy of at least 0.1% FSR
Must provide a voltage measurement precision of at least 0.02% FSR
Must provide a voltage measurement accuracy of at least 0.05% FSR
Should provide a voltage measurement accuracy of at least 0.02% FSR
Must have a temperature measurement precision of at least 0.1°C
Must have a temperature measurement accuracy of at least 2°C
Should have a temperature measurement accuracy of at least 1°C
Control accuracy
Must provide a current control accuracy of at least 5% FSR
Must provide a current control accuracy of at least 0.05% below 10 A
Should provide a current control accuracy of at least 0.1% FSR
Must provide a voltage control accuracy of at least 0.02% FSR
Must provide a power control accuracy of at least 1% FSR
Software and interfacing
Must include all hardware and software required to start using the instrument
Must have a good, clear, easy to use, and efficient graphic user interface (GUI)
Must have the software option to configure custom current profiles and power profiles that are at least 24 hours long
Must support stacking of at least 30 such profiles to run sequentially
Should be able to read measurement values and control the current setting programatically with a communication protocol (such as MODBUS, Ethernet I/P or CAN )
Notes
1. For all requirements of precision and accuracy we define full scale range (FSR) to be the range calculated between zero and the maximum positive value
2. All range, accuracy, am precision requirements are specified for the individual channels
Price
Price shall be submitted in SEK excluding VAT and include all related costs requested in this request for tender.
Agreement terms
The supplier must accept all agreement terms.
Tender submission
The tender must be received no later than 2022-06-20
The tender shall be valid until 2022-07-22
Evaluation of tenders
The most economically advantageous tender (i.e. an assessment of price, quality, service etc.) will be chosen.
Description of the criteria that will be assessed (in the prioritized order):
1. Quality of the GUI
2. Should be able to read measurement values and control the current setting programatically with a communication protocol (such as MODBUS, Ethernet I/P or CAN)
3. Should be able to provide a maximum charge and discharge current of at least 25 A
4. Should provide a current control accuracy of at least 0.1% FSR
5. Should of all measurement accuracy and precision
6. Should be able to provide a total bidirectional power flow of at least 125 W (25 A @ 5V)
7. Should have a measurement acquisition frequency of at least 15 Hz
8. Price and Should include electrical impedance spectroscopy (EIS) on each channel
Confidentiality and the freedom of information principle
After the procurement has been terminated the principle of public access to official records applies. This means that all documents related to the procurement process become public and may be disclosed to anyone who requests access to them. Exceptions may only be made if there is any specific reason to believe that the Tenderer may suffer damage if the information is disclosed.
If you have information that you feel is of such nature, your tender shall include precise details of which information should remain confidential, and how you would suffer damage in case of the public disclosure of this information. Chalmers will evaluate your request and determine whether confidentiality in accordance with Swedish law may be deemed to apply.
(2022-06-20)
Direktupphandling
2022-05-23
Direktupphandling
Västra Götalands län
F 2022-0055